  【Abstract】 Objective To appraise the protective effect with using the protective machine.Methods To have a health check for people who are engaged in interventional therapy.Results Distortion rate of chromosome has been dropping year after year.Sum total of white blood cell,white blood cell classification counts and blood platelet are in-side the range of normal condition.It is not abnormal to check crystalline lens.Conclusion The protective machine which is designed by ourselves can drop radiation for people who are engaged in interventional therapy.It can prolong working time.It can also protect people's health.It makes marked effect.

  【Key words】 interventional therapy;radiation protection;effect

  介入放射治疗作为一门新兴的临床学科,是微创医学的重要组成部分。它在X射线的介导下,采用经皮穿刺和导管技术进行诊断和治疗,具有曝光量大,多角度投照、接触时间长、近台操作等特点,给工作人员的身体健康造成极大影响。为在介入放射诊疗实践中,加强对放射工作人员的辐射防护,保障工作人员的身体健康,我们于1999年底研制了“乔丹尼”C壁X线机移动式多位相介入放射防护体 [1] ,2000年后应用于C臂X线机介入诊疗工作,现将应用5年后防护效果报告如下。

  1 材料与方法

  1.1 血象指标 按《全国临床检验操作规程》 [2] 进行。

  1.2 外周血淋巴细胞微核率(MNF)检测 取10名工作人员静脉血1.5ml,无菌肝素抗凝,1000r/min离心,吸出血浆1ml和白细胞层0.2~0.3ml加入到RPMI1640培养液4ml中,再加入植物血凝素300μg/ml,置于37℃恒温培养72h,最后制片染色和观察计数 [3] 。

  1.3 眼晶状体 使用YZ-5LSII裂隙灯检查。

  1.4 空气照射量率 在9寸视野、3D射线强度的条件下,贵州省放射卫生防护所、黔南州职业病防治所应用FJ-347A型X、γ射线巡测仪、F-3013数字γ辐射仪,进行空气照射量率测定,每个部位测定2次,取其均值。

  1.5 健康监护 贵州省黔南州疾病预防控制中心承担(原黔南州职业病防治所)。

  2 结果

  2.1 工作人员各部位所受空气照射量率 介入治疗过程中,在使用和未使用移动式多位相介入放射防护体的条件下,对操作者头、胸、手、下肢等位置测定空气照射量率,使用介入放射防护体后,操作人员所受的照射剂量达到国家规定的限值标准,见表1。

来顶一下
返回首页
返回首页
 
推荐资讯
掌间筑未来,西门子医疗推出全新CT产品SOMATOM go.
掌间筑未来,西门子医疗
“智”动创新 共享“健康中国” 西门子医疗亮相2017 HOSPEQ
“智”动创新 共享“
西门子医疗携一系列创新成果亮相CMEF 2017   助力“智慧”医疗服务建设
西门子医疗携一系列创
西门子医疗中国亮相第77届中国国际医疗器械春季博览会,多方位展示互联网时代下 “智慧”成果
西门子医疗中国亮相第
最后更新
热门点击
  1. 过度医疗之忧:被植入体内的“GDP”
  2. 卵巢囊肿疗法——超声波微电极介入疗法!
  3. 旋转DSA技术应用在肺癌支气管动脉造影中
  4. 微创介入治疗颈椎病的临床观察
  5. 介入治疗肝癌的副作用
  6. 介入治疗辐射防护效果观察
  7. 经桡动脉介入治疗进展
  8. 经桡动脉介入治疗的现状与展望
  9. 胃镜下介入治疗的临床应用分析
  10. 举办“第八届全国CT、MRI、X线影像诊断及新